Cost-Effective H₂O₂ Hydrogen Peroxide low-Temperature Plasma Sterilizer
- Point of origin
- China CE
- Throughput
- Detection mode: 21 minutes; standard mode: 35 minutes
- Application scenarios
- Hospital Endoscopy Center Operating Room
MJQ-30 Hydrogen peroxide low-temperature plasma sterilizer
★The total design volume of the sterilization chamber is ≥42.9L, and the solvent used is 28-30L.
Sterilization chamber: rectangular (square); the inner chamber is made of aviation aluminum, with a thickness of ≥8mm.
Sealed door: single door, made of aviation aluminum, with a thickness of ≥20mm.
Filling method: cassette type; each cassette has ≥12 capsules, and the hydrogen peroxide capacity of a single capsule is ≤2ml.
Capsule counting memory function: After the cassette is installed, the number of capsules used is automatically calculated, and the number of remaining capsules and the number of full cycles that can be run are prompted.
Automatic identification system: Automatically identify whether the cassette model matches the equipment, and automatically identify the validity period of the cassette.
★The concentration of hydrogen peroxide sterilant is ≤59.5%, and has the qualification for dangerous chemical business license
Amount of sterilant: detection mode: ≤2mL/time, standard mode: ≤4mL/time.
★The evaporator is equipped with a concentration purification device, and the concentration of hydrogen peroxide after purification is ≥97%
Sterilization program and sterilization time: detection mode ≤23 minutes, standard mode ≤35 minutes, and with program countdown function.
★Control system: PLC
Display: ≥7-inch TFT true color touch screen.
Recording method: micro printer prints records.
Data storage function: can store sterilization process data for at least 3 years or 30,000 cycles, which can be retrieved and printed at any time.
Can automatically monitor plasma operating power, pressure data of each stage, time of each stage, and temperature of the entire operation.
Protection device: equipped with power protector, with the function of automatically monitoring whether the user's power supply is overvoltage, undervoltage, or phase sequence error; equipped with automatic fault detection, with fault code display alarm, fault audio alarm and fault recording function; the sterilizer sealed door and sterilization chamber are equipped with temperature protection devices respectively, which can automatically perform preheating control; equipped with automatic obstacle avoidance protection function for sealing door closing.
Information interface: can be connected to the CSSD traceability system.
Pipeline: SUS304 stainless steel or higher specification pipe.
Vacuum pump: adopts a rotary vane vacuum pump resistant to hydrogen peroxide corrosion and equipped with a vacuum pump phase sequence protector.
Convenient operation: the cavity storage basket can be adjusted to 1~2 layers or more.
Biological monitoring bacteria: thermophilic Bacillus spores and Bacillus subtilis black spores.
★ It can sterilize single-channel stainless steel catheters with an inner diameter of ≥0.7mm and a length of ≤600mm or single-channel polytetrafluoroethylene catheters with an inner diameter of ≥1mm and a length of ≤4000mm.
★ The residual hydrogen peroxide in the polytetrafluoroethylene lumen after sterilization is less than 0.0008mg/cm2, and the residual hydrogen peroxide in the stainless steel lumen is less than 0.0008mg/cm2.
The evaluation of the compatibility of metal and non-metal instruments should lead to the conclusion that there is no corrosion and no impact on the expected clinical service life.
★ The residual value of hydrogen peroxide in the air for 8 hours is ≤0.6mg/m3.
★ There is a test report on the accuracy of the "Tianzhihang Orthopedic Surgery Positioning and Navigation Tool Kit" that can still achieve normal use after sterilization 125 times.
CE certification