Fully Automatic Coagulation Analyzer PM-C200T
- Point of origin
- China CE ISO13485
- Testing item
- PT,APTT,TT, FIB,DD,FDP,AT-III
- Throughput
- 200t/h
- Application scenarios
- ER,ICU,Operating Room,Warfarin Clinic,clinical lab
- Sample
- Whole blood Serum Plasma
- System type
- Open and closed system optional
The PM-C200T fully automatic coagulation analyzer is highly versatile and suitable for various departments including ER, ICU, Operating Room, Warfarin Clinic, and clinical laboratory settings.
Product Parameters
Display
Screen: 10.4-inch LCD color screen
Touch Operation: Supports glove touch screen operation
Detection Methods
Methods: Coagulation method, immunoturbidimetry, chromogenic substrate method
Detection Items
Tests: PT, APTT, TT, FIB, DD, FDP, AT-III, PC, PS, various factor analyses
Test Speed
PT: 180 tests per hour (180T/H)
DD: 20 tests per hour (20T/H)
Test Positions
Total Test Positions: 12, all controlled at 37℃ constant temperature
Coagulation Method: 8 positions, test wavelength 660nm
Chromogenic Substrate Method: 2 positions, test wavelength 405nm
Immunoturbidimetry: 2 positions, test wavelength 575nm
Incubation Positions
Incubation Positions: 12, controlled at 37℃ constant temperature
Reagent Positions
Total Reagent Positions: 23, with <15℃ refrigeration function
Cooling Efficiency: Reaches <15℃ within half an hour of startup
Sample Positions
Regular Sample Positions: 24, pull-out type, support original tube loading
Emergency Sample Positions: 8, pull-out type, support original tube loading
Refrigerated Sample Positions: 8, fixed, <15℃ refrigeration function
Reaction Cup Capacity
Total Capacity: 88 reaction cups (2 compartments, 44 cups each)
Data Storage
Storage Capacity: Can store 10,000 coagulation reaction curves and query 10,000 test results
Automated Detection System
Wavelengths: 405nm, 575nm, 660nm
Light Source: LED cold light source with long life, no need for replacement
Detection Algorithms: Algorithm and quantitative method for fibrinogen concentration
Sample Needle: Automatic heating, liquid level sensing, and anti-collision function
Alarms: Automatic alarms for insufficient reagents, reaction cups, and cleaning fluids
Sample Management
Total Sample Positions: 32 (24 regular, 8 emergency)
Rack Type: Single-row pull-out, supports emergency priority testing
Tube Support: Standard test tubes, original blood collection tubes, and 1.5mL micro sample cups
Reagent Management
Total Reagent Positions: 23, all with refrigeration function
Rack Type: Basket-type, easy to operate
Calibration: Automatic card recognition for calibration curves
Barcode Reader: Built-in in reagent compartment, automatic entry of reagent information
Software Management
Interface: Graphical all-Chinese software interface, simple and efficient
Data Storage: Can store 10,000 coagulation reaction curves
System Maintenance: Built-in system log database and maintenance interface