Lab Biological Safety Cabinet LC Series
- Point of origin
- China CE ISO13485 ISO9001
- Application scenarios
- CDC, pharmaceutical companies, university research
The LC Series Biological Safety Cabinet represents the new generation of safety cabinets designed to provide a high level of protection for both users and the environment. It combines advanced technology, user-friendly features, and stringent safety standards to ensure optimal performance in laboratory settings.
Key Features and Advantages
Compliance with International Standards:
Independently tested to meet NSF49 (US Standard), EN12469 (European Standard), and CFDA certification, ensuring compliance with global safety standards.
Large LCD Screen:
Mounted at eye level for ease of use, the durable LCD provides a comprehensive view of airflow, work area temperature, total run time, operating parameters, and real-time alarm information from Ukal Focs.
Compact Design:
Lower cabinet height (2130/2200mm) makes it suitable for laboratories with lower ceiling heights.
ULPA Filter:
Utilizes a high-efficiency ULPA filter made from silicate glass fiber from a world-renowned brand, providing over 99.9995% typical efficiency in trapping particles.
High-Quality Self-Induced Blower:
Global high-quality blower ensures more power savings, longer life, and higher stability.
Automatic compensation function maintains the exhaust air volume even when the airflow pressure of the filter drops by 50%.
Airflow Sensor:
Two independent sensors for downflow and exhaust airflow provide double security and real-time display of airflow speed.
Biosafety Protection Tests:
Passed rigorous pressure decay tests.
KI-Discus and microbial tests ensure the safety of the environment, personnel, and samples.
Filter Life Indication:
Filter life indicator measures filter life based on actual conditions.
Provides visual and acoustic alarms when the filter needs attention.
Ultraviolet Lamp with Timing Function:
Allows for setting the opening and closing times of the ultraviolet lamp for enhanced sterilization control.
Active Carbon Postfilters:
Real-time filter life display with visual and audio reminders for maintenance.
Energy-Saving Operation Mode:
Features an energy-saving function with low-speed operation when the glass door is closed, reducing energy consumption.
| Model | 900Lc | 1200Lc | 1500 Lc | 1800 Lc |
| Level of BSC | Class II Type A2 | |||
| Airflow Mode | 70% cycle,30% exhaust | |||
| Average Airflow Velocity(Downflow) | 0.35m/s(68.90fpm) | |||
| Average Airflow Velocity(Inflow) | 0.53m/s(104.33fpm) | |||
| Personnel Protection Test | KI-Discus containment and microbiological testing are performed | |||
| Product Protection Test 1-8x10^6(threetimesinsuession | ≤5 CFU | |||
| Cross-contamination Test 1-8x10^6(three times in succession) | ≤2 CFU | |||
| Filter Type | ULPA | |||
| Filter Typical Efficiency | Filters provide 99.9995% typical efficiency for particle size greater than 0.12 microns | |||
| Airflow Volume(Downflow) | ≈718m3/h | ≈945m2/h | ≈1172m3/h | ≈1280m3/h |
| Airflow Volume(exhaust) | ||||
| RMS | ≤3μm | |||
| Fluorescent Light Intensity | >1000lux | |||
| Sound Emission | ≤60dBA | ≤63dBA | ≤65dBA | ≤65dBA |
| Maximum power | 1600W | 1600W | 2000W | 2200W |
| Power supply | AC220±10%,50HZ | |||
| External Dimensions with Base Stand(WxDxH) | 1040x790x2130mm | 1340x790x2130mm | 1640x790x2130mm | 1940x790x2130mm |
| Internal Work Area, Dimensions(WxDxH) | 950x575x625mm | 1250x575x625mm | 1550x575x625mm | 1850x575x625mm |